Page 76 CMD300 CONTROL SOFTWARE
C
YBER
A
MP
380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC.
27)
RAW
The RAW command allows the user to pass an arbitrary command to the CyberAmp. The
remainder of the command line is passed unedited to the CyberAmp. CMD300 will add the
leading "AT" command prefix and device number if applicable. Any text response from the
CyberAmp is displayed.
28)
READ
PROBE <probe number> <probe field>
This command reads and displays information stored in the EEPROM of any SmartProbe
attached to the CyberAmp. The <probe number> is any allowed channel number for this
CyberAmp. The <probe field> can be one of the keywords:
ALL, MODEL, SERIAL, NAME, MANUFACTUREDATE, CALIBRATEDATE,
INPUTCOUPLING, FILTER, UNITS, SCALEFACTOR or ZEROVALUE
29)
REPEAT <repeat count>
REPEAT instructs CMD300 to repeat the following commands the number of times specified.
If <repeat count> is omitted then the following commands are repeated indefinitely until the
<Esc> key is hit. This command is most useful when used with the DELAY command. The
following command monitors the overload status of all the channels by continuously polling
the CyberAmp.
CMD300 REPEAT OVERLOAD
30)
RESCAN
The CMD300 program keeps track of the maximal channel number of each CyberAmp
connected to a serial port via the Axon expandable RS232 bus. This maximal channel
number is preserved in the .INI file between runs of CMD300. If you need to switch between
CyberAmp 320 and CyberAmp 380 units on the same serial port using the same Device
Number then you must erase the .INI file or use the RESCAN command to have the CMD300
program re-poll the CyberAmp to determine the maximal channel number supported by the
current CyberAmp unit. This command should only be needed during final checkout by Axon
test technicians.
31)
RESTART
The RESTART Command is used only in the command-file mode. It restarts execution of the
command file, beginning at line 1.
32)
SERIAL
The SERIAL command should only be used during final checkout of the CyberAmp by Axon
test technicians. It is used to set the EEPROM based serial number of the CyberAmp unit.
The user is prompted with the current serial number and is asked to confirm that the serial
number should be changed.