Microsoft LPF-00004 Manual do Operador Página 40

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 39
Page 22 GENERAL INFORMATION
CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC.
Expansion
To increase the number of signal conditioning channels, if your software supports it, up to ten
CyberAmp 380s may be "daisy chained" to one computer serial port. The RS-232 OUT port of the
first CyberAmp is connected to the RS-232 IN port of the next CyberAmp, and so on. The
interconnection cable should be a simple 25-pin flat cable, such as the RS232-02 cable from Axon
Instruments.
Each CyberAmp or compatible instrument on the daisy-chained serial bus must be set to a unique
device address by using the ten-position rotary switch on the back panel. It is not necessary for the
first instrument on the bus to receive address number 1. Device addresses can be allocated randomly
as long as the addresses are unique.
Vista de página 39
1 2 ... 35 36 37 38 39 40 41 42 43 44 45 ... 156 157

Comentários a estes Manuais

Sem comentários