GENERAL INFORMATION Page 19
CYBERAMP 380, COPYRIGHT MARCH 2002, AXON INSTRUMENTS, INC.
REFERENCE SECTION — GENERAL INFORMATION
Power Supply Voltage Selection and Fuse Changing
Supply Voltage
The CyberAmp 380 operates from all international supply voltages. The two input ranges are:
1) 115 V : For 100 V
ac
to 125 V
ac
operation.
2) 230 V : For 200 V
ac
to 250 V
ac
operation.
Note: When shipped, the AC mains fuse and fuse holder are removed from the back of the
CyberAmp 380 and placed in a bag stapled to the handle. All instruments shipped
from Axon Instruments are set for a supply voltage of 115 V
ac
. If the supply voltage
is
not
115 V
ac
, have a qualified electronics technician perform the supply voltage
change procedure according to the following instructions.
To change the supply voltage setting:
1)
Disconnect the power cord
2) For 115 V operation -- slide the switch on the back panel towards the label "115".
For 230 V operation -- slide the switch on the back panel towards the label "230".
3) Reconnect the power cord.
Changing the Fuse
The CyberAmp 380 uses an 0.5 A, 250 V, slow acting, 5 x 20 mm fuse on both voltage ranges.
Before changing the fuse investigate the reason for its failure.
To change the fuse:
1)
Disconnect the power cord.
2) Use a screwdriver or a similar device to rotate the fuse holder counterclockwise.
3) Replace the fuse with another fuse of the same rating.
4) Reconnect the power cord.